Cardinal Health recalls 15Fr JP-2188 channel drains due to performance issues like lack of radiopacity and broken tips. Affected units have specific lot numbers and UDI/DI codes.
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.
Cardinal Health is recalling 15Fr Jackson-Pratt Channel Drains with model JP-2188 because they have performance issues such as lack of radiopacity, dull trocars, and broken tips. These issues could lead to improper wound drainage and potential complications during medical procedures.
The drains lack radiopacity, making them harder to see on X-rays. Dull trocars and broken tips can cause improper drainage or damage to the wound site during medical procedures.
Medical professionals using these drains for wound drainage procedures are most at risk. Patients receiving these drains may experience improper drainage or complications due to the product issues.
Check for lot numbers N210278 to N220163 and UDI/DI codes 50885380182431 or 20885380182430 on the product packaging.
Look for lot numbers N210278 to N220163 and UDI/DI codes 50885380182431 or 20885380182430 on the packaging.
Cardinal Health received complaints about performance issues including lack of radiopacity, dull trocars, lack of flexibility, and broken tips.
Check the product for lot numbers N210278 to N220163 and UDI/DI codes 50885380182431 or 20885380182430. If affected, contact Cardinal Health for further instructions.
The product issues could lead to improper drainage or wound complications, but the recall is for performance issues rather than immediate safety hazards.
We’ve drafted a message you can send Cardinal Health to request your refund or repair — edit it as you like.
Subject: Recall Z-0969-2023 — Cardinal Health Cardinal Health Hello, I own a Cardinal Health that is covered by recall Z-0969-2023 from Cardinal Health. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.