Cardinal Health recalls 10Fr JP-2221 channel drains due to performance issues like lack of radiopacity and broken tips. Affected units have specific lot numbers and UDI/DI codes.
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.
Cardinal Health is recalling 10Fr Jackson-Pratt Channel Drains with model JP-2221 because they have performance issues such as lack of radiopacity, dull trocars, and broken tips. This could lead to improper wound drainage and potential complications during medical procedures.
The channel drains may not show up clearly on X-rays (lack of radiopacity), have dull trocars that affect function, and break easily, which could lead to improper wound drainage and complications during medical procedures.
Medical professionals and patients using the recalled channel drains for wound drainage procedures. Those with the specific lot numbers or UDI/DI codes listed are most at risk.
Check for the model number JP-2221, lot numbers N210278 through N220163, and UDI/DI codes 50885380182554(cs) or 20885380182553(ea).
Identify if your channel drain has the model JP-2221 and the listed lot numbers or UDI/DI codes.
Cardinal Health received complaints about performance issues including lack of radiopacity, dull trocars, lack of flexibility, and broken tips.
Check if your channel drain has the model JP-2221 and the specific lot numbers or UDI/DI codes listed in the recall notice.
The product may not show up clearly on X-rays and could break, potentially leading to improper wound drainage during medical procedures.
We’ve drafted a message you can send Cardinal Health to request your refund or repair — edit it as you like.
Subject: Recall Z-0981-2023 — Cardinal Health Cardinal Health Hello, I own a Cardinal Health that is covered by recall Z-0981-2023 from Cardinal Health. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.