Cardinal Health recalls 10mm Channel Drains (JP-2212) due to performance issues like lack of radiopacity and broken tips. Affected units have specific lot numbers and UDI/DI codes.
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.
Cardinal Health is recalling 10mm Channel Drains (JP-2212) because they have been reported to lack radiopacity, have dull trocars, lack flexibility, and break easily. This could lead to improper wound drainage and complications during medical procedures.
The drains can become dull, lack flexibility, and break easily, which may cause improper drainage of wounds. This could lead to infection or delayed healing if the drain doesn't function correctly.
Medical professionals using these drains for wound drainage procedures are most at risk. Patients receiving these drains during surgery or post-op care may experience complications if the drains malfunction.
Check for the model number JP-2212 and lot numbers N210278 through N220163. The UDI/DI codes 50885380182493 (cs) and 20885380182492 (ea) also identify affected units.
Identify the lot numbers and UDI/DI codes listed in the recall notice to confirm if your drain is affected.
Cardinal Health received complaints about performance issues including lack of radiopacity, dull trocars, lack of flexibility, and broken tips.
Check the lot numbers and UDI/DI codes to see if your drain is affected; the recall does not require immediate action but should be addressed.
The drains may not function properly, potentially leading to infection or delayed healing if they break or lack flexibility.
We’ve drafted a message you can send Cardinal Health to request your refund or repair — edit it as you like.
Subject: Recall Z-0975-2023 — Cardinal Health Cardinal Health Hello, I own a Cardinal Health that is covered by recall Z-0975-2023 from Cardinal Health. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.