CareFusion recalls Alaris PC Units 8015 with specific software and wireless network cards that may fail to connect, risking infusion data transfer issues. Affected units have IP addresses from x.x.x.224 to x.x.x.255.
Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed
CareFusion is recalling Alaris PC Units 8015 that use specific wireless network cards and software versions. These units may fail to connect to hospital systems, potentially requiring manual infusion programming. This could disrupt critical patient data transfers in medical settings.
The PC units may experience network connection failures, preventing them from receiving new data or transferring patient information to hospital systems. This could force healthcare staff to manually reprogram infusions, increasing the risk of errors during critical patient care.
Healthcare facilities using Alaris PC Units 8015 with affected software versions and wireless network cards are most at risk. Patients receiving infusions via these systems may face delays in data transfer if the unit fails to communicate with hospital systems.
Check for Alaris PC Unit 8015 models with software versions 12.1.0, 12.1.2, or 12.1.3 using wireless network cards that support 802.11 b/g or 802.11 a/b/g. Affected units have IP addresses from x.x.x.224 to x.x.x.255.
Verify the software version on your Alaris PC Unit 8015; affected versions are 12.1.0, 12.1.2, or 12.1.3.
Ensure your wireless network card supports 802.11 b/g or 802.11 a/b/g, as these are the affected card types.
Check if your unit's IP address falls within the range x.x.x.224 to x.x.x.255.
No, but you should check for affected software versions and network cards to ensure safe operation.
The unit may require manual programming of infusions, which could delay patient care if not handled carefully.
Check your software version (12.1.0, 12.1.2, or 12.1.3), network card type, and IP address range as described in the recall.
We’ve drafted a message you can send CareFusion to request your refund or repair — edit it as you like.
Subject: Recall Z-1192-2023 — CareFusion CareFusion Hello, I own a CareFusion that is covered by recall Z-1192-2023 from CareFusion. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.