Omnicell recalls IVX Station units with model 298921373 that may prepare incorrect doses due to scale inaccuracies. Affected units have specific serial numbers. Contact Omnicell for remedy.
IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling
Omnicell is recalling IVX Station devices used for preparing pharmaceuticals because they might make doses that are too high or too low. This could lead to unsafe medication levels for patients.
The IVX Station may not properly account for variations in the amount of medication being prepared, leading to doses that are outside the labeled accuracy range. This could result in medication being too strong or too weak for patients.
Healthcare facilities using the Omnicell IVX Station for preparing medications are most affected. Patients receiving medications made with these devices could be at risk if doses are inaccurate.
Check for model number 298921373 on the device. Affected units have serial numbers starting with IVXS000014 through IVXS000019.
Look for model number 298921373 on the IVX Station device.
Reach out to Omnicell directly for instructions on how to proceed with the recall.
The affected model number is 298921373.
Serial numbers starting with IVXS000014 through IVXS000019 are affected.
Check for model number 298921373 and serial numbers starting with IVXS000014 through IVXS000019.
We’ve drafted a message you can send Omnicell to request your refund or repair — edit it as you like.
Subject: Recall Z-1617-2023 — Omnicell Omnicell Hello, I own a Omnicell that is covered by recall Z-1617-2023 from Omnicell. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.