Percussionaire recalls Distal Phasitron Part S20020 for use with VDR-4 control units due to potential assembly error causing less effective therapy. Affected units have specific UDI-DI and expiration dates.
Breathing circuit reservoir bag may be assembled at the wrong attachment point, which will result in less than optimal therapy.
Percussionaire is recalling a specific part (S20020) for their VDR-4 control units because it might be assembled incorrectly, leading to less effective therapy. This could mean the device doesn't work as well as it should, but it doesn't pose an immediate safety risk.
The breathing circuit reservoir bag might be attached at the wrong point during assembly. This can cause the therapy to be less effective than intended, but it does not cause injury or malfunction.
Users of Percussionaire VDR-4 control units with the Distal Phasitron Part S20020 are affected. Those who have the specific UDI-DI and expiration dates listed are most at risk.
Check for the Distal Phasitron Part S20020 with UDI-DI 00849436000693 or 00849436000694, and expiration dates September 13, 2024 or November 14, 2024.
Verify the UDI-DI and expiration dates on the part to confirm if it's affected.
No, the recall is for potential reduced therapy effectiveness, not a safety hazard.
The official recall does not specify a remedy, so contact Percussionaire Corporation directly for guidance.
Look for UDI-DI 00849436000693 with expiration September 13, 2024, or UDI-DI 00849436000694 with expiration November 14, 2024.
We’ve drafted a message you can send Percussionaire to request your refund or repair — edit it as you like.
Subject: Recall Z-0693-2024 — Percussionaire Percussionaire Hello, I own a Percussionaire that is covered by recall Z-0693-2024 from Percussionaire. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.