Draeger Medical recalls VentStar Basic (P)250 ventilators due to risk of breathing circuit detachment from loose connections during ventilation. All units are affected.
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Draeger Medical is recalling VentStar Basic (P)250 ventilators because glued connections can become loose, potentially causing parts to detach during use. This could lead to breathing issues if the circuit becomes disconnected.
The breathing circuit connections can become loose over time, especially during ventilation. If pieces detach, it could interrupt the patient's breathing and cause breathing difficulties.
Healthcare providers using VentStar Basic (P)250 ventilators are most at risk. Patients receiving ventilation through this device could be affected if the circuit detaches.
Check for VentStar Basic (P)250 ventilators with model number MP00352. All units are affected regardless of lot or serial number.
Inspect the breathing circuit connections for looseness before each use.
Reach out to Draeger Medical for further guidance on securing connections.
Yes, all VentStar Basic (P)250 units with model number MP00352 are affected.
Look for the model number MP00352 on your VentStar Basic (P)250 ventilator. All units are affected regardless of lot or serial number.
No, but you should check connections before each use and contact Draeger Medical for further guidance.
We’ve drafted a message you can send Draeger Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-1441-2023 — Draeger Medical Draeger Medical Hello, I own a Draeger Medical that is covered by recall Z-1441-2023 from Draeger Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.