NOXBOX recalls specific nitric oxide delivery devices due to potential check valve misalignment causing gas leaks or therapy interruptions. Affected units have unique serial numbers listed in the recall notice.
Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a worst case scenario, there may be an interruption in therapy which could lead to oxygen desaturation of the patient and associated complications.
NOXBOX is recalling certain nitric oxide delivery systems that could malfunction due to a check valve issue. This might cause oxygen or nitric oxide leaks, or interrupt therapy, risking oxygen desaturation for patients.
The device's check valve can misalign, causing leaks of nitric oxide or oxygen gas. If the device fails, it might not deliver therapy properly, leading to oxygen desaturation and health complications.
Patients using the NOXBOXi Nitric Oxide Delivery System for medical therapy are most at risk. The recall affects specific units with unique serial numbers listed in the notice.
Check your device's serial number against the list of affected units provided in the recall notice. The affected units have specific serial numbers starting with NI followed by numbers.
Compare your device's serial number to the list of affected serial numbers provided in the recall notice.
The device may experience malfunctions due to misalignment of the check valve in the manifold, which could cause nitric oxide leaks, oxygen gas leaks, or interruption in therapy.
Check your device's serial number against the list of affected serial numbers provided in the recall notice.
The recall notice does not specify a remedy, so contact NOXBOX directly for further instructions.
We’ve drafted a message you can send NOXBOX to request your refund or repair — edit it as you like.
Subject: Recall Z-1667-2023 — NOXBOX NOXBOX Hello, I own a NOXBOX that is covered by recall Z-1667-2023 from NOXBOX. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.