Philips recalls Tempus LS Manual Defibrillators with part number 00-3020 that may show "HW failure DPM" error, causing ineffective pacing. Affected units have serial numbers before 7022.001634.
Defibrillator/pacemaker module (DPM) may encounter a problem causing communication from the Mainboard to the DPM to fail, leading to no or ineffective pacing with displayed error message "HW failure DPM".
Philips is recalling Tempus LS Manual Defibrillators with part number 00-3020 that may show an error message indicating ineffective pacing. This could lead to serious health risks if the device fails during use.
The defibrillator's communication between the mainboard and pacing module may fail, causing the device to show an error message "HW failure DPM". This could result in ineffective pacing, potentially leading to serious health issues during critical cardiac events.
Medical professionals using the defibrillator for cardiac care are most at risk. Patients relying on the device for life support could face complications if the error occurs.
Check the defibrillator's part number, which is 00-3020. Affected units have serial numbers starting with digits before 7022.001634.
Look for the error message "HW failure DPM" on the defibrillator display.
Confirm the serial number starts with digits before 7022.001634.
Part number 00-3020.
Check the serial number starts with digits before 7022.001634.
The defibrillator may display "HW failure DPM".
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-1825-2023 — Philips Philips Hello, I own a Philips that is covered by recall Z-1825-2023 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.