Medline recalls specific ultrasound probe cover kits with potential inadequate seam barriers. These kits may pose a risk during diagnostic ultrasound procedures. Affected models were distributed from December 2017 to May 2023.
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Medline is recalling certain ultrasound probe cover kits that may have an inadequate barrier at the seams. This could lead to contamination during diagnostic ultrasound procedures if the covers are used improperly.
The probe covers may have an inadequate barrier at the seams, potentially allowing contaminants to enter the ultrasound probe during procedures. This could lead to contamination of the probe and increase infection risk for patients.
Healthcare professionals using these kits for diagnostic ultrasound procedures are most at risk. Patients receiving these procedures may also be affected if the covers are compromised.
Check for model numbers DYNDH1112, DYNDH1165A, or DYNDH1204 on the kit. These kits were distributed by Medline between December 2017 and May 2023.
Look for model numbers DYNDH1112, DYNDH1165A, or DYNDH1204 on the kit packaging.
The probe covers may have an inadequate barrier at the seams, potentially allowing contaminants to enter the ultrasound probe during procedures.
Models DYNDH1112, DYNDH1165A, and DYNDH1204 were distributed between December 2017 and May 2023.
No, but healthcare professionals should check model numbers and follow the recall instructions if applicable.
We’ve drafted a message you can send Medline to request your refund or repair — edit it as you like.
Subject: Recall Z-2198-2023 — Medline Medline Hello, I own a Medline that is covered by recall Z-2198-2023 from Medline. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.