Wilson-Cook recalls Hemospray Endoscopic Hemostats with part numbers HEMO-10 and HEMO-10-EU due to risk of endoscope sticking to tissue during use.
There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope.
Wilson-Cook is recalling Hemospray Endoscopic Hemostats that may cause the endoscope to stick to tissue, making it difficult or impossible to remove. This could lead to complications during medical procedures.
The Hemospray powder can stick to the endoscope's tip, causing it to adhere to tissue. This makes it hard or impossible to move the endoscope during surgery, potentially causing complications.
Medical professionals using the Hemospray Endoscopic Hemostat for procedures are most at risk. Patients undergoing procedures with this device may also be affected.
Check for part numbers HEMO-10 (REF G21049) or HEMO- 10-EU (REF G21346) on the device. The device was sold with unexpired lot numbers.
Look for part numbers HEMO-10 (REF G21049) or HEMO-10-EU (REF G21346) on the device.
Yes, the Hemospray powder may stick to the endoscope, causing it to adhere to tissue and making it difficult to remove during procedures.
The recall record does not specify a remedy, so contact Wilson-Cook Medical for further instructions.
Check for part numbers HEMO-10 (REF G21049) or HEMO-10-EU (REF G21346) on the device.
We’ve drafted a message you can send Wilson-Cook Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-1979-2023 — Wilson-Cook Medical Wilson-Cook Medical Hello, I own a Wilson-Cook Medical that is covered by recall Z-1979-2023 from Wilson-Cook Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.