Scican recalls STATCLAVE G4 tabletop autoclaves for possible incomplete sterilization of non-steam-compatible items. Affected units may cause damage or unsafe sterilization if used improperly.
The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and / or damage to the autoclave.
Scican is recalling its STATCLAVE G4 tabletop autoclaves because they can't safely sterilize liquid loads or materials not compatible with steam. Using them for these items risks incomplete sterilization or damaging the appliance.
The STATCLAVE G4 is designed only for sterilizing medical instruments validated for steam use. It cannot handle liquid loads or materials that might damage the device. Using it for incompatible items risks incomplete sterilization or appliance damage.
Healthcare providers who own or use STATCLAVE G4 autoclaves for medical sterilization are most at risk. Patients and staff could be exposed to unsafe medical instruments if the device fails to properly sterilize items.
Check for model number G4 622303 or serial numbers starting with 710523E00005-00007. These units were sold with the UDI-DI (01)07540196001793.
Look for model number G4 622303 or serial numbers starting with 710523E00005-00007 on the device.
No, the STATCLAVE G4 is designed only for sterilizing medical instruments validated for steam use. It cannot safely handle liquid loads.
Using it for items not compatible with steam sterilization may cause incomplete sterilization or damage to the autoclave.
Affected units have model number G4 622303 and serial numbers starting with 710523E00005-00007.
We’ve drafted a message you can send Scican to request your refund or repair — edit it as you like.
Subject: Recall Z-2286-2023 — Scican Scican Hello, I own a Scican that is covered by recall Z-2286-2023 from Scican. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.