Covidien recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheters with potential occlusion from excessive silicone lubricant. Affected units may cause treatment delays or complications.
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Covidien is recalling certain MAHURKAR dialysis catheters due to a risk of blockage from excessive silicone lubricant. This could delay treatment or cause complications like reduced blood flow or clotting.
The catheter tip can become blocked by silicone lubricant (MDX) that coats it. This blockage may cause treatment delays, reduced blood flow, or clotting if the lubricant dislodges from the catheter.
Patients receiving dialysis and healthcare providers using the catheters are most at risk. The specific units affected are those with the listed lot numbers.
Check for the model: MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, IC Tray. The lot numbers 2133600127, 2133700049, and 2200400180 are affected.
Identify the model number and lot numbers to confirm if your catheter is affected.
Reach out to your healthcare provider for guidance on next steps if you have an affected catheter.
The catheter tip can become blocked by excessive silicone lubricant (MDX), which may cause treatment delays, reduced blood flow, or clotting.
Lot numbers 2133600127, 2133700049, and 2200400180 are affected.
Check for the model: MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, IC Tray with the listed lot numbers.
We’ve drafted a message you can send Covidien to request your refund or repair — edit it as you like.
Subject: Recall Z-2367-2023 — Covidien Covidien Hello, I own a Covidien that is covered by recall Z-2367-2023 from Covidien. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.