Carefusion recalls specific lots of surgical grasper devices that may crack from improper adhesive use during manufacturing. Affected units have lot numbers B23 or C23. Contact Carefusion for details.
Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Carefusion is recalling specific surgical grasper devices that could crack due to a manufacturing defect involving adhesive. This defect might cause the device to break during use, potentially leading to injury if the device fails unexpectedly.
The grasper devices may crack or break because of a manufacturing defect where adhesive was applied incorrectly. This could cause the device to fail during surgical procedures, potentially leading to injury if the device breaks unexpectedly.
Surgical professionals using the specific lot numbers B23 or C23 of the Carefusion grasper devices. Patients and staff at medical facilities using these devices are at risk if the device breaks during surgery.
Check for lot numbers B23 or C23 on the device's packaging or label. The product is a surgical grasper used in endoscopic procedures.
Look for lot numbers B23 or C23 on the device's packaging or label to confirm if it's affected.
The graspers may crack or break due to incorrect adhesive use during manufacturing.
Check for lot numbers B23 or C23 on the device's packaging or label.
Contact Carefusion directly for further instructions on how to handle the affected device.
We’ve drafted a message you can send Carefusion to request your refund or repair — edit it as you like.
Subject: Recall Z-2428-2023 — Carefusion Carefusion Hello, I own a Carefusion that is covered by recall Z-2428-2023 from Carefusion. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.