Olympus recalls two bronchovideoscope models due to risk of becoming lodged in endotracheal tubes during medical procedures, potentially causing airway blockage.
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Olympus is recalling two bronchovideoscope models (BF-3C40 and BF-N20) because they might get stuck in endotracheal tubes during medical procedures. This could block the airway and cause serious breathing problems for patients.
The bronchovideoscope's diameter is too large for the endotracheal tube connector, which can cause it to become lodged during medical procedures. This may block the airway and prevent proper breathing for patients.
Healthcare professionals using these scopes during medical procedures are most at risk. Patients receiving care with these scopes could also be affected if the scope becomes lodged.
Check for the model numbers BF-3C40 or BF-N20 on the device. The product was sold with specific serial numbers and UDI-DI codes listed in the recall notice.
Look for the model numbers BF-3C40 or BF-N20 on the bronchovideoscope.
The bronchovideoscope models might become lodged in the endotracheal tube connector due to their diameter being too large, which could block the airway.
Healthcare providers should check if they have the recalled models and contact Olympus for guidance on next steps.
The recall record does not mention a specific fix or remedy for this issue.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-2502-2023 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-2502-2023 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.