3M recalls specific lots of Attest Super Rapid 5 Steam-Plus Steam Challenge Packs that may have biological indicators with melting caps after steam sterilization at 132°C or 135°C. Affected units could fail during sterilization cycles.
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.
3M is recalling specific lots of Attest Super Rapid 5 Steam-Plus Steam Challenge Packs because some units may have biological indicators with caps that melt, deform, or crack after steam sterilization at 132°C or 135°C. This could cause the sterilization process to fail, potentially leading to contamination of medical equipment.
The biological indicators in these packs have caps that may melt, deform, or crack when exposed to steam sterilization cycles at 132°C or 135°C. This could cause the indicators to fail during sterilization, meaning medical equipment might not be properly sterilized and could become contaminated.
Healthcare facilities using these specific lots of 3M Attest Super Rapid 5 Steam-Plus Steam Challenge Packs for steam sterilization processes are most at risk. The affected units are typically used in medical settings to monitor sterilization effectiveness.
Check for the lot codes 41482VF (expiring 33P4ND) or 41482VF (expiring 01/02/2025) on the product packaging. These packs are designed for steam sterilization processes in healthcare settings.
Look for lot codes 41482VF with expiration dates 33P4ND or 01/02/2025 on the packaging.
The biological indicators in some lots may have caps that melt, deform, or crack after steam sterilization at 132°C or 135°C, potentially causing sterilization failure.
Check for lot codes 41482VF with expiration dates 33P4ND or 01/02/2025 on the product packaging.
No, but you should check your product lot codes and follow the action steps to ensure safe use.
We’ve drafted a message you can send 3M to request your refund or repair — edit it as you like.
Subject: Recall Z-2640-2023 — 3M 3M Hello, I own a 3M that is covered by recall Z-2640-2023 from 3M. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.