Stryker recalls disposable Spetzler-Malis bipolar forceps with incorrect 54-month expiration date label. Products should be used within 36 months of manufacture to avoid potential quality issues.
An error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
Stryker is recalling disposable Spetzler-Malis bipolar forceps with an incorrect expiration date label. The product shows a 54-month shelf life but should only be used for 36 months. This could lead to potential quality issues if used beyond the correct shelf life.
The product label incorrectly states a 54-month shelf life when the actual shelf life is 36 months. Using these instruments beyond the correct shelf life could lead to potential quality issues, though serious injury is not expected.
Medical professionals who use these disposable surgical instruments are likely affected. Patients using these instruments during procedures may be at risk if the instruments are used beyond the correct shelf life.
Check for the product name 'Disposable Spetzler-Malis Bipolar forceps, 23CM IM DISP BAYONET 1.0MM TIP' with catalog number 6770-230-010. Lot numbers 200507015, 200515010, and 201008014 are also affected.
Verify the expiration date on the product label matches the 36-month shelf life from the manufacturer.
The product label incorrectly states a 54-month shelf life when the actual shelf life is 36 months.
Check the expiration date on the product label to ensure it is within the 36-month shelf life. If the label shows a longer shelf life, contact Stryker for guidance.
No serious injury is expected from using these instruments within the correct shelf life. The recall is for potential quality issues if used beyond the 36-month period.
We’ve drafted a message you can send Stryker to request your refund or repair — edit it as you like.
Subject: Recall Z-0251-2024 — Stryker Stryker Hello, I own a Stryker that is covered by recall Z-0251-2024 from Stryker. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.