Olympus recalls EUS scopes with potential water contamination causing infection risk. Affected models include GF-UC140P-AL5, GF-UCT180, GF-UE160-AL5. Contact for repair or return.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Olympus is recalling certain endoscopic scopes that may have water in the channel after repair, which could lead to infection. This is a safety issue for medical users handling these devices.
Water remaining in the channel after repair can cause contamination. This contamination may lead to infections during medical procedures if not properly addressed.
Medical professionals using Olympus EUS scopes with model numbers GF-UC140P-AL5, GF-UCT180, or GF-UE160-AL5 are affected. Patients undergoing procedures with these scopes may be at risk of infection.
Check for model numbers GF-UC140P-AL5, GF-UCT180, or GF-UE160-AL5. These scopes were sold with specific serial numbers listed in the recall notice.
Reach out to Olympus directly for instructions on how to return or repair the affected scopes.
The recall states that water in the channel after repair could lead to infection risk, so affected scopes should not be used until repaired or returned.
Model numbers GF-UC140P-AL5, GF-UCT180, and GF-UE160-AL5 are affected.
Check the model number on your scope against the listed numbers: GF-UC140P-AL5, GF-UCT180, or GF-UE160-AL5.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-0333-2024 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-0333-2024 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.