BioFire recalls Torch Base (HTFA-ASY-0104) due to increased risk of power switch degradation causing potential short circuits or failure. Affected units have serial numbers listed in the recall notice.
Due to an increased risk of degradation of the power entry switch resulting from arcing inside of the power switch that may result in carbon build-up on the switch contacts. The carbon build-up may lead to excess heating inside of the power switch, subsequently leading to deformation of the power switch case. The deformation can create an opportunity for a power switch electrical short. In addition to an electrical short, deformation may result in an open circuit causing the power switch to fail. This event would only occur after the product is in use.
BioFire is recalling its Torch Base (HTFA-ASY-0104) because a power switch can degrade over time, leading to overheating, short circuits, or complete failure. This risk occurs only after the device is in use and may affect lab equipment used for diagnostic testing.
The power switch inside the device can degrade due to arcing, causing carbon buildup. This buildup may lead to overheating, deformation of the switch case, and potential electrical shorts or complete failure of the power switch. This issue only occurs after the device is used regularly.
Users of the BioFire Torch Base (HTFA-ASY-0104) for diagnostic testing, particularly in clinical or lab settings. Those with the specific serial numbers listed in the recall notice are most at risk.
Check if your BioFire Torch Base has the model number HTFA-ASY-0104 and the serial numbers listed in the recall notice (e.g., KTB00006 through KTB02971).
Identify if your device has a serial number from the list provided in the recall notice (e.g., KTB00006 through KTB02971).
If your device is affected, stop using it immediately to prevent potential electrical issues.
The recalled device is the BioFire Torch Base with model number HTFA-ASY-0104.
If your device has a serial number from the list provided in the recall notice, stop using it immediately and contact BioFire for further instructions.
No, this recall specifically affects the BioFire Torch Base (HTFA-ASY-0104) with the listed serial numbers.
We’ve drafted a message you can send BioFire to request your refund or repair — edit it as you like.
Subject: Recall Z-0353-2024 — BioFire BioFire Hello, I own a BioFire that is covered by recall Z-0353-2024 from BioFire. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.