C.R. Bard recalls Sensica Urine Output System (SCCS1002) due to incorrect time displays when connected to WiFi, causing inaccurate monitoring data. Affected units have specific serial numbers listed in the recall.
BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed during a patient monitoring session which resulted in 1) incorrect time displayed on the monitoring screen, 2) incorrect hour block timestamps on the monitoring screen, 3) incorrect hour timestamps on data screens, and 4) incorrect catheter use time.
C.R. Bard is recalling the Sensica Urine Output System (SCCS1002) because connecting it to WiFi can cause incorrect time displays on monitoring screens. This leads to inaccurate patient monitoring data, which could affect medical decisions.
When the device connects to WiFi, it displays incorrect times during patient monitoring sessions. This causes errors in the hour blocks and catheter use time recorded in the system, potentially leading to inaccurate medical data.
Patients using the Sensica Urine Output System for medical monitoring are affected. Healthcare providers who use the device with WiFi connectivity are most at risk of inaccurate data.
Check for the catalog number SCCS1002 on the device. Affected units have specific serial numbers listed in the recall notice, including BMGPRMYS001, BMGPRMYS003, and other serial numbers provided in the recall.
Look for the serial number on the device to see if it matches the list of affected serial numbers provided in the recall.
C.R. Bard has received user complaints that when the device connects to WiFi, it displays incorrect times during patient monitoring sessions, leading to inaccurate hour blocks and catheter use time.
The affected devices have catalog number SCCS1002.
Check the device's serial number against the list of affected serial numbers provided in the recall notice.
We’ve drafted a message you can send C.R. Bard to request your refund or repair — edit it as you like.
Subject: Recall Z-0554-2024 — C.R. Bard C.R. Bard Hello, I own a C.R. Bard that is covered by recall Z-0554-2024 from C.R. Bard. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.