Howmedica recalls SCORPIO TS Tib Insert (part 72-4-7512) due to potential oxidation in units over 5 years old, which may affect material properties.
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Howmedica is recalling SCORPIO TS Tib Insert knee replacement parts that are over 5 years old because they may have elevated oxidation levels in the UHMWPE material. This could affect the part's performance and safety.
The UHMWPE material in the part can oxidize over time, especially in units older than 5 years. Oxidation may weaken the material, potentially affecting its performance during knee replacement procedures.
Patients who received this knee replacement part from Howmedica after 2019 are likely affected. Those with older units (over 5 years) face the highest risk.
Check for the part number 72-4-7512 and lot number 5M6LM1 on the product. This recall applies to units manufactured over 5 years ago.
Look for part number 72-4-7512 and lot number 5M6LM1 on the product.
The product may have elevated oxidation levels in the UHMWPE material after 5 years of age, which could affect its material properties.
Check for part number 72-4-7512 and lot number 5M6LM1 on the product. This recall applies to units manufactured over 5 years ago.
The recall does not specify a remedy, so contact Howmedica directly for guidance.
We’ve drafted a message you can send Howmedica to request your refund or repair — edit it as you like.
Subject: Recall Z-0877-2024 — Howmedica Howmedica Hello, I own a Howmedica that is covered by recall Z-0877-2024 from Howmedica. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.