Zyn Medical recalls Z800WF infusion pumps with software version 5.2.05 that may revert alarm volumes, risking therapy delays or organ failure. Affected units were sold to healthcare professionals.
When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in therapy and in extreme cases, associated risk of organ failure or death
Zyn Medical is recalling Z800WF infusion pumps with software version 5.2.05 because a software bug can cause alarm volumes to reset to low settings, potentially delaying critical therapy and causing organ failure or death. Healthcare professionals using these pumps should stop using them immediately.
The software bug causes the alarm volume to unexpectedly switch from a higher setting to a low one when using the patient query feature. This can delay critical therapy, and in extreme cases, lead to organ failure or death.
Healthcare professionals who own or use Z800WF pumps with software version 5.2.05 are most at risk. These pumps are intended for intravenous therapy under medical supervision.
Check for the model Z-800WF with software version 5.2.05. These pumps were sold to healthcare professionals for intravenous therapy.
Immediately stop using Z800WF pumps with software version 5.2.05 as they may cause therapy delays or organ failure.
The software bug causes alarm volumes to reset to low settings when using the patient query feature, risking therapy delays and potential organ failure or death.
Check for the model Z-800WF with software version 5.2.05. These pumps were sold to healthcare professionals for intravenous therapy.
Stop using the pump immediately. The manufacturer does not provide a remedy for this specific software issue.
We’ve drafted a message you can send Zyn: Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-1183-2024 — Zyn: Medical Zyn: Medical Hello, I own a Zyn: Medical that is covered by recall Z-1183-2024 from Zyn: Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.