GE Healthcare recalls Lullaby Incubators (models XP, TR, standard) due to bedside panel potentially not latching even when appearing closed, risking patient falls if disengaged.
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
GE Healthcare is recalling three types of neonatal incubators because the bedside panel might not latch properly even when it looks closed. If a patient hits an unlatched panel or porthole, it could fall open and cause the baby to fall.
The bedside panel can appear closed but not be securely latched if the red indicator light is visible. If a baby bumps into this unlatched panel or porthole, the panel might open and cause the baby to fall.
Newborns in hospitals using GE Healthcare Lullaby Incubators are most at risk. Parents or caregivers should check if their incubator model is listed in the recall.
Check for model numbers: Lullaby Incubator (M1151042, QCR031), Lullaby XP (8320, M1151135, QCR034), or Lullaby TR (M1151357). All units with these models are affected.
Look for the red indicator light; if visible, the panel may not be latched even when it appears closed.
Confirm your incubator model matches the recalled types: Lullaby, Lullaby XP, or Lullaby TR with specific model numbers listed.
The bedside panel can appear closed but not be securely latched if the red indicator light is visible, risking patient falls if the panel disengages.
Lullaby Incubator (model numbers M1151042, QCR031), Lullaby XP (8320, M1151135, QCR034), and Lullaby TR (M1151357).
Check the model numbers on your incubator; if they match the listed Lullaby models (XP, TR, or standard), it is affected.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0958-2024 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-0958-2024 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.