GE Healthcare recalls Care Plus hood door kits for potential panel disengagement if not latched, risking patient falls. Affected units include specific incubators and part numbers.
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
GE Healthcare is recalling hood door kits for certain incubators because the panel may appear closed but not be securely latched. If a patient hits an unlatched panel, it could fall open and cause injury.
The bedside panel can look closed but not be latched when the red indicator is visible. If a patient impacts an unlatched panel or porthole, the panel may disengage and fall open, potentially causing falls.
Parents or caregivers of infants using the affected incubators are most at risk. The recall applies to specific models and part numbers sold globally.
Check for part numbers: 6600-0082-850, 6600-0504-800, 6600-0148-800, or 6600-0970-500. The recall covers all serial numbers of the Care Plus incubators listed.
Look for the red indicator to determine if the panel is latched properly.
If the panel appears closed but the red indicator is visible, it may not be latched and should be checked for safety.
The panel may appear closed but not be latched when the red indicator is visible, risking patient falls if impacted.
Care Plus Model 1000, 2000, 3000, 4000 Incubators, Lullaby Incubators, Lullaby TR Incubators, and Lullaby XP Incubators.
Part numbers: 6600-0082-850, 6600-0504-800, 6600-0148-800, and 6600-0970-500.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0960-2024 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-0960-2024 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.