Stratasys recalls TrueDent Clear TDM 100 resin cartridges with expired dates due to formatting errors causing system to accept expired materials. Affected lots 22994 (exp. 02-Apr-2024) and 23425 (exp. 07-Nov-2024).
Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
Stratasys is recalling TrueDent Clear TDM 100 dental resin cartridges because the expiration date format error causes the system to accept expired materials. This could lead to using expired resin, which may not meet quality standards for dental appliances.
The system incorrectly switches the month and day in expiration dates, so it may not recognize expired resin. This means users could continue using resin that has expired, potentially affecting the quality of dental appliances.
Dental professionals using TrueDent Clear TDM 100 resin cartridges with specific lot numbers. Those most at risk are users who rely on the system to prevent expired resin usage.
Check for lot numbers 22994 (expires 02-Apr-2024) or 23425 (expires 07-Nov-2024) on the TrueDent Clear TDM 100 resin cartridges. The product is sold under STRATASYS LTD.
Look for lot numbers 22994 or 23425 on the resin cartridges.
The system has a formatting error that switches the month and day in expiration dates, causing it to accept expired resin.
Check the lot number on the cartridge; if it's 22994 or 23425, contact Stratasys for guidance.
The recall record does not mention a fix or remedy for the system error.
We’ve drafted a message you can send Stratasys to request your refund or repair — edit it as you like.
Subject: Recall Z-2090-2025 — Stratasys Stratasys Hello, I own a Stratasys that is covered by recall Z-2090-2025 from Stratasys. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.