Raye's recalls SW Bari Lift & Transfer Model 38060000 due to potential failure under weight limits and risk of falls for people with disabilities during use. All units with serial numbers 00845699001040 are affected.
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
Raye's is recalling its SW Bari Lift & Transfer model 38060000 because it may not safely support the maximum weight specified and lacks a safety feature to prevent falls for people with disabilities. This could lead to serious injury if used improperly.
The lift may not handle the maximum weight it's designed for when a safety factor of two times the load is applied. It also lacks a safety mechanism to prevent falls if the lifting device fails, which is especially risky for people with disabilities.
People who own or use the SW Bari Lift & Transfer model 38060000 are affected. Those with disabilities using the device are at higher risk of falling during lifting operations.
Check for the model number 38060000 on the device. All units with serial numbers starting with 00845699001040 are affected.
Look for the serial number on the device, which starts with 00845699001040.
The product may not safely support maximum weight and lacks a fall prevention feature for people with disabilities, so it should not be used until a fix is made.
Check the serial number on the device; affected units have serial numbers starting with 00845699001040.
The recall does not specify a remedy, so contact Raye's Inc. for guidance on next steps.
We’ve drafted a message you can send Raye's to request your refund or repair — edit it as you like.
Subject: Recall Z-1595-2024 — Raye's Raye's Hello, I own a Raye's that is covered by recall Z-1595-2024 from Raye's. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.