Cook recalls Yueh Centesis Disposable Catheter Needles due to potential packaging failure that could compromise sterility. Affected units have specific lot numbers and reference part numbers. Consumers should contact Cook for details.
Packaging may have low seal strength and not meet peel strength specifications, compromising device sterility
Cook is recalling Yueh Centesis Disposable Catheter Needles because their packaging might not seal properly, risking contamination. This could lead to infections if the needles are used without proper sterility.
The packaging may have low seal strength, which could allow contamination if the needle is not sterile when used. This might happen during the manufacturing process or storage.
Medical professionals and patients who use these needles for procedures like paracentesis or thoracentesis. Those with the specific lot numbers or reference part numbers are most at risk.
Check for the reference part number DTVN-5.0-19-10.0 and order number G09490 on the packaging. Lot numbers 15787864 and 15804398 are also associated with this recall.
Look for the reference part number DTVN-5.0-19-10.0 and order number G09490 on the packaging.
The Yueh Centesis Disposable Catheter Needle with reference part number DTVN-5.0-19-10.0 and order number G09490.
Check for the reference part number DTVN-5.0-19-10.0, order number G09490, and lot numbers 15787864 or 15804398 on the packaging.
The recall does not specify a remedy, so contact Cook directly for further instructions.
We’ve drafted a message you can send Cook to request your refund or repair — edit it as you like.
Subject: Recall Z-1527-2024 — Cook Cook Hello, I own a Cook that is covered by recall Z-1527-2024 from Cook. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.