Beckman Coulter recalls DxI 9000 Access Immunoassay Analyzers with cracks in the wash buffer tank that may leak, causing test delays and exposure risks. Affected units have specific serial numbers.
Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.
Beckman Coulter is recalling certain DxI 9000 Access Immunoassay Analyzers due to cracks in the wash buffer tank that can leak. This may cause delays in patient test results, exposure to hazards, and equipment damage. Affected units have specific serial numbers.
Cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank can leak. This may cause delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.
Healthcare facilities using the DxI 9000 analyzer with specific serial numbers are most at risk. The recall affects laboratory equipment used for patient testing.
Check for serial numbers: 300254, 300256, 300257, 300258, 300259, 300263, 300264, 300266, 300268, 300269, 300270, 300271, 300272, 300273, 300275, 300281, and 300282. The device is a DxI 9000 Access Immunoassay Analyzer.
Inspect the bottom of the Internal Wash Buffer Tank for cracks around the threaded insert.
Beckman Coulter is recalling certain DxI 9000 Access Immunoassay Analyzers due to cracks in the wash buffer tank that can leak.
The affected units have serial numbers: 300254, 300256, 300257, 300258, 300259, 300263, 300264, 300266, 300268, 300269, 300270, 300271, 300272, 300273, 300275, 300281, and 300282.
Cracks in the wash buffer tank can leak, causing delays in patient test results, exposure to physical and chemical hazards, and potential equipment damage.
We’ve drafted a message you can send Beckman Coulter to request your refund or repair — edit it as you like.
Subject: Recall Z-1546-2024 — Beckman Coulter Beckman Coulter Hello, I own a Beckman Coulter that is covered by recall Z-1546-2024 from Beckman Coulter. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.