Electro Medical Systems SA recalls WATER BOTTLE model EG-121 due to risk of disassembly under pressure during dental treatments. Affected bottles may separate when used or stored without releasing pressure.
Received complaints that bottles assembled with one batch of bottle collars can be disassembled in two parts when used under pressure (during treatment or storage if the use is not releasing the pressure inside the bottles).
Electro Medical Systems SA is recalling WATER BOTTLE model EG-121 because some bottles can disassemble into two parts when used under pressure during dental treatments. This could happen if the bottle is not properly released after use or stored without releasing pressure.
The bottles can disassemble into two parts when used under pressure during dental treatments or stored without releasing the pressure inside. This disassembly could cause injury if the parts separate unexpectedly.
Dentists and dental patients who use WATER BOTTLE model EG-121 for dental treatments are affected. Those who store the bottle without releasing pressure may be at higher risk.
Check for model number EG-121 with UDI code 07613353180287 and lot numbers 16457-10 through 16866-80. The bottles are sold empty and filled by users for dental treatments.
Inspect the bottle for signs of disassembly or unusual pressure buildup before use.
Ensure the bottle is properly released after use to prevent disassembly during storage.
The recall is due to complaints that some bottles assembled with one batch of bottle collars can disassemble into two parts when used under pressure during dental treatments or stored without releasing pressure.
Check for model number EG-121 with UDI code 07613353180287 and lot numbers 16457-10 through 16866-80. The bottles are sold empty and filled by users for dental treatments.
The recall does not include a specific remedy; users should check the bottle condition and use with care to prevent disassembly.
We’ve drafted a message you can send WATER BOTTLE to request your refund or repair — edit it as you like.
Subject: Recall Z-1687-2024 — WATER BOTTLE WATER BOTTLE Hello, I own a WATER BOTTLE that is covered by recall Z-1687-2024 from WATER BOTTLE. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.