Philips Respironics recalls all Trilogy Evo Universal Ventilators except software version 1.05.06.00 due to potential sudden loss of ventilation from battery or power alarms.
Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Philips Respironics is recalling all Trilogy Evo Universal Ventilators except those with software version 1.05.06.00. The device may lose ventilation suddenly if battery or power alarms trigger when the unit still has power.
The ventilator may show a battery or power alarm even when it still has sufficient power. This could cause sudden loss of ventilation if the alarm goes off without the user realizing the unit is still working.
Pediatric and adult patients using the Trilogy Evo Universal Ventilator. Patients with respiratory conditions requiring ventilator support are most at risk.
Check for the model number DS2000X11B on the device. The ventilator was sold with all software versions except 1.05.06.00.
Look for the software version on the device; if it's not 1.05.06.00, it's affected.
The ventilator may show a battery or power alarm when it still has power, which could cause sudden loss of ventilation. It is not safe to use without checking the software version.
Check the software version; if it's not 1.05.06.00, contact Philips Respironics for instructions.
The affected ventilators have model number DS2000X11B and software versions other than 1.05.06.00.
We’ve drafted a message you can send Philips Respironics to request your refund or repair — edit it as you like.
Subject: Recall Z-1505-2024 — Philips Respironics Philips Respironics Hello, I own a Philips Respironics that is covered by recall Z-1505-2024 from Philips Respironics. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.