Recalled American Contract Systems laparoscopy kits may have compromised sterility due to insufficient aeration time during sterilization. Affected kits have specific UDI-DI and lot numbers. Consumers should contact the manufacturer for details.
Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility
American Contract Systems is recalling laparoscopy convenience kits that may not have been properly sterilized because the aeration time was too short. This could lead to infections if the kits are used without proper sterility.
The kits may have been sterilized improperly because the aeration time was too short. This could compromise sterility, potentially leading to infections if the kits are used without proper sterilization.
Healthcare professionals using laparoscopy kits for surgical procedures are most at risk. The recall affects specific kits with the UDI-DI 00191072190536 and Lot 981241.
Check for the UDI-DI number 00191072190536 and lot number 981241 on the kit packaging. These kits were sold with an expiration date of January 19, 2025.
Reach out to American Contract Systems for specific details about the recalled kits and any necessary actions.
The lot number is 981241.
The expiration date is January 19, 2025.
Look for the UDI-DI number 00191072190536 and lot number 981241 on the kit packaging.
We’ve drafted a message you can send American Contract Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-2519-2024 — American Contract Systems American Contract Systems Hello, I own a American Contract Systems that is covered by recall Z-2519-2024 from American Contract Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.