Canon Medical recalls Vantage Orian, Elan, Titan and Galan MRI systems due to potential fire hazard from cable disconnection. Affected models include MRT-1550, MRT-2020, MRT-1504 and MRT-3010/3020 with specific serial numbers.
The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
Canon Medical is recalling several MRI systems that could cause fire if the power cable disconnects. This happens because the terminal block securing the cable wasn't designed properly, which might lead to arcing and melting.
The terminal block that holds the power cable for the MRI's gradient coil wasn't secured as intended. This could cause the cable to disconnect, leading to electrical arcing, melting of the cable, smoke, and fire.
Medical facilities using the recalled MRI systems are affected. Hospitals and clinics that own or operate these specific models are most at risk.
Check if your MRI system has the model numbers: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), or Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020). The specific serial numbers listed in the recall notice indicate affected units.
The recalled MRI systems pose a fire risk if the power cable disconnects. Do not operate these units until the recall remedy is completed.
The affected models are Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), and Vantage Galan 3T MRT-3020 (MEXL-3020). Each has specific serial numbers listed in the recall notice.
Check your system's model number and serial number against the list provided in the recall notice. Affected units have the model numbers and serial numbers listed in the description.
Stop using the system immediately and contact Canon Medical to arrange for a replacement or repair. The recall requires immediate action to prevent fire risk.
We’ve drafted a message you can send Canon Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-2041-2024 — Canon Medical Canon Medical Hello, I own a Canon Medical that is covered by recall Z-2041-2024 from Canon Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.