Megadyne recalls MEGA SOFT reusable patient return electrodes for children under 12 due to burn risk. Use only for patients 12+ to prevent injury.
To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.
Megadyne is recalling its MEGA SOFT reusable patient return electrodes because they can cause burns in children under 12. The product is designed for patients 12 years and older to reduce this risk.
The electrodes can cause burns if used on patients younger than 12. This is because the product is designed for patients weighing over 25 pounds (11.4kg) and aged 12 or older.
Patients under 12 years old are at risk of burns from these electrodes. Adults using the product for children under 12 should stop use immediately.
Look for the product name 'MEGADYNE MEGA SOFT Reusable Patient Return Electrode' with model number 0830. The product is sold for patients weighing over 25lbs (11.4kg).
Ensure the patient is 12 years or older before using the electrode. Do not use on children under 12.
The recall is to limit use to patients aged 12 or older to reduce the risk of burns.
The official notice does not mention a fix; the remedy is to use only for patients 12 or older.
The product has model number 0830 and is labeled for patients weighing over 25lbs (11.4kg).
We’ve drafted a message you can send Megadyne to request your refund or repair — edit it as you like.
Subject: Recall Z-2249-2024 — Megadyne Megadyne Hello, I own a Megadyne that is covered by recall Z-2249-2024 from Megadyne. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.