B. Braun recalls Infusomat 60DROP METRISET Pump Sets (lots after Aug 10, 2:02 PM, 2023) due to potential backflow of medication causing patient injuries. Affected units may need replacement.
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
B. Braun is recalling Infusomat 60DROP METRISET Pump Sets made after August 10, 2023, because a faulty backcheck valve could cause medication to flow backward into the main IV line, risking patient injury. This could lead to adverse drug reactions or loss of medication.
The backcheck valve might malfunction, allowing medication from a secondary IV container to flow back into the primary IV container. This could cause medication to be lost or lead to adverse drug reactions in patients.
Healthcare professionals using the Infusomat pump set for IV medication administration are most at risk. Patients receiving IV treatments with this specific pump set could be affected.
Check for the Infusomat 60DROP METRISET Pump Set model number 362034, distributed after August 10, 2023. The product is used with electrically-powered IV pumps for medication administration.
Identify the model number (362034) and distribution date (after August 10, 2023) to confirm if your pump set is affected.
The recall indicates a potential risk of backflow, so affected units should be checked for the specific model and distribution date before use.
The recall does not specify a remedy, so contact B. Braun Medical for guidance on whether to stop using or replace the pump set.
Check for model number 362034 and distribution date after August 10, 2023. The product is used with electrically-powered IV pumps for medication administration.
We’ve drafted a message you can send B. Braun Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-2380-2024 — B. Braun Medical B. Braun Medical Hello, I own a B. Braun Medical that is covered by recall Z-2380-2024 from B. Braun Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.