PeGa Medical recalls FRT250 4.0-6.4 cartridges due to incorrect component used during assembly. Affected units include specific lot numbers from the Fassier-Duval Telescopic IM System.
During the assembly of the FRT250 cartridge assembly, an incorrect component was used. Specifically, the FRT240-HEX (2) was used instead of the FRT250-HEX.
PeGa Medical is recalling FRT250 4.0-6.4 cartridges because the wrong component was used during assembly. This could lead to improper function of the telescopic IM system, but the recall does not specify immediate safety risks.
The recall states that an incorrect component (FRT240-HEX) was used instead of the correct FRT250-HEX during assembly. This mismatch could cause the cartridge to function improperly in the telescopic IM system.
Users of the Fassier-Duval Telescopic IM System with the FRT250 cartridge (lot numbers F281-04 or F462-02) are affected. Medical professionals using this specific component are most at risk.
Check if your FRT250 4.0-6.4 cartridge has lot numbers F281-04 or F462-02. The cartridge is part of the Fassier-Duval Telescopic IM System.
Identify the lot number on your FRT250 cartridge to see if it matches F281-04 or F462-02.
The recall states that an incorrect component (FRT240-HEX) was used instead of the correct FRT250-HEX during assembly.
The recall does not specify a remedy, so check the lot number and contact PeGa Medical for further instructions.
The recall does not mention any safety risks or injuries, so the risk is low.
We’ve drafted a message you can send PeGa Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-2941-2024 — PeGa Medical PeGa Medical Hello, I own a PeGa Medical that is covered by recall Z-2941-2024 from PeGa Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.