Alcon recalls specific ClearCut Safety Slit SB ophthalmic knives due to reported sharpness issues. Affected models include 8065772745 with lot numbers 1668AM through 16FXK8. Consumers should check their product details and contact Alcon for guidance.
Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.
Alcon is recalling certain ClearCut Safety Slit SB ophthalmic knives (model 8065772745) because they have been reported to be too sharp, which could cause injury during use. This recall affects ophthalmic surgeons and medical professionals who use these knives for eye procedures.
The knives have been reported to be too sharp, which could lead to injury if used improperly. This sharpness issue may cause cuts or other injuries during eye surgery procedures.
Ophthalmic surgeons and medical professionals who use the ClearCut Safety Slit SB ophthalmic knives for eye procedures are most at risk. The recall affects specific model numbers and lot numbers listed in the product details.
Check for the model number 8065772745 on the product. The affected lot numbers range from 1668AM to 16FXK8. This product is an ophthalmic knife used in eye surgery.
Look for the model number 8065772745 and lot numbers 1668AM through 16FXK8 on the product.
The specific model number is 8065772745.
The affected lot numbers range from 1668AM to 16FXK8.
Check for the model number 8065772745 and lot numbers 1668AM through 16FXK8 on the product.
We’ve drafted a message you can send Alcon to request your refund or repair — edit it as you like.
Subject: Recall Z-2654-2024 — Alcon Alcon Hello, I own a Alcon that is covered by recall Z-2654-2024 from Alcon. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.