Beaver Visitec recalls CustomEyes Procedure Packs with Sol-M Luer Lock and Luer Slip 1ml syringes (part 58001825) due to potential air embolism risk. Affected units sold with specific batch numbers.
Due to a voluntary recall of low dead space syringes, and luer slip tip syringes manufactured by Sol-Millennium Medical (Sol-M), Inc. These syringes manufactured by Sol-M are present in some of BVI CustomEyes Procedure Packs.
Beaver Visitec is recalling CustomEyes Procedure Packs that contain Sol-M Luer Lock and Luer Slip 1ml syringes. These syringes may have a low dead space design that could cause air embolism if used improperly.
The syringes have a low dead space design that may allow air to enter the bloodstream during injections, potentially causing air embolism. This risk is higher when the syringes are used for procedures where air can be drawn into the bloodstream.
Medical professionals using CustomEyes Procedure Packs with the specified syringes are most at risk. Patients receiving procedures with these syringes could also be affected.
Check for the part number 58001825 on the syringe packaging. Affected packs were sold with batch numbers starting with 6074001.
Look for part number 58001825 on the syringe packaging or label.
The recalled packs contain Sol-M Luer Lock and Luer Slip 1ml syringes with part number 58001825.
Check for part number 58001825 on the syringe packaging or batch number 6074001 on the label.
The recall does not specify a remedy, so contact the manufacturer for guidance.
We’ve drafted a message you can send Beaver Visitec to request your refund or repair — edit it as you like.
Subject: Recall Z-2917-2024 — Beaver Visitec Beaver Visitec Hello, I own a Beaver Visitec that is covered by recall Z-2917-2024 from Beaver Visitec. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.