Nova Ortho-Med recalls Monarch Rollator Red walkers with model 4329RD due to fall risk when transitioning from slouched to upright seating.
Firm identified a potential risk for a fall hazard. When the end user is readjusting from the slouched seated position to an upright seated position, there is the potential for the movement to cause the backrest to lift out of the backrest holder and disengage from the unit.
Nova Ortho-Med is recalling Monarch Rollator Red walkers (model 4329RD) because they can disengage from the backrest holder during seat adjustments, potentially causing falls. This hazard affects users who move from a slouched seated position to an upright position.
The backrest can lift out of the backrest holder when the user moves from a slouched seated position to an upright seated position. This disengagement could cause the walker to become unstable, increasing the risk of a fall.
Users of the Monarch Rollator Red walker model 4329RD are most at risk when transitioning from a slouched seated position to an upright seated position. This includes individuals who use walkers for mobility assistance.
Check for the model number 4329RD on the walker. The serial number range is HT4329RDME0001 to HT4329RDMM0420. This product was sold with the UPC code 652308012263.
Inspect the backrest holder to ensure it stays securely attached when moving from a slouched to upright seated position.
Model 4329RD with serial numbers from HT4329RDME0001 to HT4329RDMM0420.
The backrest may lift out of the backrest holder when transitioning from a slouched seated position to an upright seated position.
The recall does not specify a remedy, so check with Nova Ortho-Med for guidance.
We’ve drafted a message you can send Nova Ortho-Med to request your refund or repair — edit it as you like.
Subject: Recall Z-0131-2025 — Nova Ortho-Med Nova Ortho-Med Hello, I own a Nova Ortho-Med that is covered by recall Z-0131-2025 from Nova Ortho-Med. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.