Baxter Healthcare recalls TruSystem 7000 operating tables with battery short circuits and smoke emissions. Affected units have specific product codes and serial numbers. Stop using immediately to prevent fire or injury.
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
Baxter Healthcare is recalling its TruSystem 7000 operating tables due to reported battery short circuits and smoke emissions. This hazard could lead to fire or injury if the device is used without addressing the issue.
The batteries and connectors in these operating tables can short circuit and emit smoke, which may cause fire or injury. This hazard occurs during normal use when the device is powered on.
Healthcare facilities using the TruSystem 7000 operating tables with product code REF 1841082 are affected. Medical staff and patients using these tables are at risk of fire or injury.
Check for the product code REF 1841082 and serial numbers starting with 00887761974234 on the device. Affected units were sold with this specific model and code.
Discontinue use of the TruSystem 7000 operating table with product code REF 1841082 to prevent fire or injury risks.
The product code is REF 1841082.
Check for the product code REF 1841082 and serial numbers starting with 00887761974234 on the device.
The recall record does not mention any remedy or replacement options.
We’ve drafted a message you can send Baxter Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0215-2025 — Baxter Healthcare Baxter Healthcare Hello, I own a Baxter Healthcare that is covered by recall Z-0215-2025 from Baxter Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.