Smiths Medical recalls TranStar Monitoring Sets with Product Code REF MXG300260 due to potential inaccurate pressure readings. Affected units may show incorrect readings or fail to zero properly. Contact for replacement or return.
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Smiths Medical is recalling TranStar Monitoring Sets with Product Code REF MXG300260 because a manufacturing defect can cause inaccurate pressure readings or prevent the device from zeroing correctly. This could lead to incorrect blood pressure measurements, which is critical for medical monitoring.
The device may show inaccurate pressure readings or fail to zero properly due to a leak or defect in the disposable pressure transducer or its chip. This can lead to incorrect blood pressure measurements, which is critical for medical monitoring.
Healthcare professionals using the TranStar Monitoring Set for blood pressure monitoring are most at risk. Patients relying on this device for accurate readings could experience incorrect measurements.
Check for Product Code REF MXG300260 on the device. Lot Numbers 4270120, 4297963, 4316640, 4319604, 4323118, 4287727, 4359814, 4387976, 4340473 are affected.
Verify the Product Code REF MXG300260 and Lot Numbers listed in the recall notice.
Due to a manufacturing defect that can cause inaccurate pressure readings or prevent the device from zeroing correctly.
Check the Product Code and Lot Numbers to see if your device is affected. Contact Smiths Medical for instructions on replacement or return.
The recall is for potential inaccurate readings, which could affect medical monitoring, but the record does not specify a risk of serious injury.
We’ve drafted a message you can send Smiths Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-0204-2025 — Smiths Medical Smiths Medical Hello, I own a Smiths Medical that is covered by recall Z-0204-2025 from Smiths Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.