Smiths Medical recalls TranStar DPT pressure monitors with possible inaccurate readings due to manufacturing defects. Affected units may show pressure interruptions or zeroing issues. Contact for replacement or repair.
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Smiths Medical is recalling TranStar DPT pressure monitors that may give inaccurate pressure readings or stop working properly due to a manufacturing defect in the pressure transducer. This could lead to incorrect medical monitoring, which is serious for patients using the device.
The device may show inaccurate pressure readings or stop working entirely because of a leak or defect in the pressure trans:ducer chip. This can cause incorrect medical monitoring, which is critical for patient safety.
Healthcare professionals using the TranStar DPT pressure monitors for medical monitoring are most at risk. Patients relying on these devices for accurate pressure readings could be affected if the device malfunctions.
Check for the product code REF MX950SCFT, lot numbers 4277663, or the UDI/DI numbers 50351688503945 (case) and 10351688503947 (pouch).
Verify if the device shows inaccurate pressure readings or fails to zero properly.
Reach out to Smiths Medical for a replacement or repair if the device is affected.
Due to a manufacturing defect that may cause inaccurate pressure readings, interruptions in pressure monitoring, or inability to zero the device.
Check for product code REF MX950SCFT, lot number 4277663, or UDI/DI numbers 50351688503945 (case) and 10351688503947 (pouch).
Contact Smiths Medical for a replacement or repair if the device shows inaccurate readings or fails to zero properly.
We’ve drafted a message you can send Smiths Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-0193-2025 — Smiths Medical Smiths Medical Hello, I own a Smiths Medical that is covered by recall Z-0193-2025 from Smiths Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.