Becton Dickinson recalls BD EpiCenter Data Management System Version 440887 (Catalog No: 440887) due to potential unauthorized access to product service credentials, risking data confidentiality and security. Affected units include specific serial numbers listed in the recall.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Becton Dickinson is recalling BD EpiCenter Data Management System Version 440887 (Catalog Number 440887) because unauthorized actors accessed product service credentials. This could lead to breaches in data confidentiality, integrity, or availability. Users should check if their device has the listed serial numbers.
Unauthorized actors accessed product service credentials used by BD technical support teams. This could allow unauthorized access to the system, potentially compromising patient data confidentiality, system integrity, or data availability if not corrected.
Users of BD EpiCenter Data Management System Version 440887 with the specific serial numbers listed in the recall are affected. Healthcare facilities using this system for patient data management are most at risk.
Check if your device has the specific serial numbers listed in the recall (e.g., 6GF3TD2, 2MJHML1, etc.). The product version is 440887 with catalog number 440887.
Verify if your BD EpiCenter Data Management System has the specific serial numbers listed in the recall (e.g., 6GF3TD2, 2MJHML1, etc.).
If your device is affected, contact Becton Dickinson to update product service credentials to prevent unauthorized access.
The product is BD EpiCenter Data Management System Version 440887 with Catalog Number 440887.
Check if your device has a serial number listed in the recall (e.g., 6GF3TD2, 2MJHML1, etc.).
No, the recall does not require immediate discontinuation of use. Affected users should update credentials to prevent unauthorized access.
We’ve drafted a message you can send Becton Dickinson to request your refund or repair — edit it as you like.
Subject: Recall Z-0585-2025 — Becton Dickinson Becton Dickinson Hello, I own a Becton Dickinson that is covered by recall Z-0585-2025 from Becton Dickinson. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.