Medtronic recalls Becker External Drainage and Monitoring Systems with cracks or leaks in stopcocks. Affected models include EDMS 24146, EDMS II 25120, and others. Stop using if cracks or leaks are present.
Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
Medtronic is recalling Becker External Drainage and Monitoring Systems due to cracks or leaks in stopcocks. This can cause fluid leaks, potentially leading to infections or other complications if the system is used improperly.
Cracks or leaks in the stopcocks can allow cerebrospinal fluid to escape from the system. This may lead to infections, increased risk of complications, or improper monitoring of fluid pressure and flow.
Patients using the Becker External Drainage and Monitoring System for cerebrospinal fluid drainage are most at risk. Healthcare providers who use these systems for medical monitoring may also be affected.
Check for cracks or leaks in the stopcocks of the Becker External Drainage and Monitoring System. Affected models include EDMS 24146, EDMS II 25120, and others with specific lot numbers listed in the recall notice.
Inspect the stopcocks for cracks or leaks. If present, stop using the system immediately.
Affected models include EDMS 24146, EDMS II 25120, 26040 KIT, 27609, 27670, 27672, 27702, 27761, 27767, 27779, 27811, 27931, 46128, and 46129 with specific lot numbers listed in the recall notice.
Check for cracks or leaks in the stopcocks. Affected models include EDMS 24146, EDMS II 25120, and others with specific lot numbers listed in the recall notice.
If you find cracks or leaks in the stopcocks, stop using the system immediately. Contact Medtronic for further instructions.
We’ve drafted a message you can send Becker to request your refund or repair — edit it as you like.
Subject: Recall Z-0661-2025 — Becker Becker Hello, I own a Becker that is covered by recall Z-0661-2025 from Becker. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.