BD Bio Sciences recalls FACSLyric flow cytometers with specific power supply models due to potential failure causing lab staff inhalation risks and patient specimen issues. Affected units have catalog number 662876 and power supply serial numbers starting with V2309, V2318, or V2330.
On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users.
BD Bio Sciences is recalling certain FACSLyric flow cytometers that may fail to power on or stay on, risking lab staff inhalation of fumes and patient specimen loss. This affects instruments with specific power supply serial numbers starting with V2309, V2318, or V2330.
The power supply module (part number 650781) can fail, causing the instrument to stop working or remain on. If this happens during lab use, staff may inhale fumes leading to shortness of breath or coughing. Patients might need to return for additional specimen collection, risking bruising, pain, or bleeding.
Laboratory staff and patients using BD FACSLyric flow cytometers with catalog number 662876 and power supply serial numbers starting with V2309, V2318, or V2330 are most at risk.
Check for catalog number 662876 and power supply serial numbers starting with V2309, V2318, or V2330 on the instrument. The device is a flow cytometer used in labs for blood and tissue analysis.
Check the power supply serial number on the instrument to see if it starts with V2309, V2318, or V2330.
If the power supply is not affected, continue using the instrument as normal according to the manufacturer's instructions.
The faulty power supply module has part number 650781 (Power Module AC-DC 250W 12/24V M3S4S6).
No, the recall instructs customers to continue normal operation after inspecting for the specific power supply serial numbers.
If the power supply fails while in use, lab staff may inhale fumes leading to shortness of breath or coughing. Patients might need to return for additional specimen collection, risking bruising, pain, or bleeding.
Check the power supply serial number on the instrument; affected units have serial numbers starting with V2309, V2318, or V2330.
We’ve drafted a message you can send BD Bio Sciences to request your refund or repair — edit it as you like.
Subject: Recall Z-0746-2025 — BD Bio Sciences BD Bio Sciences Hello, I own a BD Bio Sciences that is covered by recall Z-0746-2025 from BD Bio Sciences. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.