Mercury Medical recalls pediatric emergency resuscitators with part numbers 1050805-1050842 due to potential failure to maintain proper pressure ranges during emergency breathing support.
Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.
Mercury Medical is recalling pediatric resuscitators that may not maintain proper pressure settings during emergency breathing support. This could lead to ineffective ventilation and harm patients needing immediate respiratory assistance.
The resuscitator's circuit flow controller has an undersized spring that can cause it to fail at the high and low ends of the pressure range. This may result in loss of positive pressure, making it ineffective for providing proper breathing support during critical emergencies.
Parents and caregivers of children who use these resuscitators for emergency respiratory support are most at risk. Healthcare providers using the device for pediatric emergencies may also be affected.
Check for part numbers 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, or 1050842 on the device. These models were sold with specific serial numbers listed in the recall notice.
Look for the part number on the device label to determine if it's affected by this recall.
Reach out to Mercury Medical directly for guidance on next steps if you have the recalled part number.
The resuscitator's circuit flow controller has an undersized spring that may cause it to fail at the high and low ends of the pressure range, potentially leading to ineffective ventilation during emergencies.
Part numbers 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, and 1050842 are affected.
The recall does not require immediate discontinuation; however, affected users should check with Mercury Medical for guidance on next steps.
We’ve drafted a message you can send Mercury Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-0821-2025 — Mercury Medical Mercury Medical Hello, I own a Mercury Medical that is covered by recall Z-0821-2025 from Mercury Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.