GE Healthcare recalls Discovery NM 630 systems with specific model numbers due to potential detector fall risk during transport or relocation without proper support. This could lead to life-threatening injury. Affected units include models H2401CF, H2401CH, H2401JE, H3100JY, and 5376204-30-2.
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
GE Healthcare is recalling certain Discovery NM 630 nuclear medicine systems that could fall during transport or relocation without proper detector support, potentially causing life-threatening injury. Owners should check if their unit has the listed model numbers to determine if it's affected.
The detector mounting mechanisms can experience excessive load during transport or relocation without proper support. This may cause the detector to fall, leading to life-threatening bodily injury.
Medical facilities using GE Healthcare Discovery NM 630 systems with model numbers H2401CF, H2401CH, H2401JE, H3100JY, or 5376204-30-2 are most at risk. The hazard occurs when the system is moved without proper detector support.
Check if your system has model numbers: 5376204-30-2, H2401CF, H2401CH, H2401JE, or H3100JY. The system may also have specific serial numbers listed in the detailed recall information.
Identify your system's model number from the list: 5376204-30-2, H2401CF, H2401CH, H2401JE, or H3100JY.
Reach out to GE Healthcare for further assistance or to confirm if your system is affected.
The detector could fall during transport or relocation without proper support, potentially causing life-threatening injury.
Check if your system has model numbers: 5376204-30-2, H2401CF, H2401CH, H2401JE, or H3100JY.
No, but you should check if your system is affected and contact GE Healthcare for guidance.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0890-2025 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-0890-2025 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.