GE Healthcare recalls Infinia II Hawkeye 4 models H2799PP, H2799PR, H2799YC, H2799YD, H3000WC and others due to potential detector fall from improper relocation. This could lead to life-threatening injury. Affected units have specific serial numbers and system IDs.
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
GE Healthcare is recalling certain Infinia II Hawkeye 4 nuclear medicine imaging systems because they could fall if moved without proper support. This might cause serious injury. Affected models include H2799PP, H2799PR, H2799YC, H2799YD, and H3000WC with specific serial numbers.
If the detector is not properly supported during transport or relocation, excessive force can be applied to the mounting mechanism. This could cause the detector to fall, potentially leading to serious injury.
Healthcare facilities that own or use these specific nuclear medicine imaging systems are most at risk. The recall applies to systems that were transported or relocated without proper detector support.
Check if your system has model numbers H2799PP, H2799PR, H2799YC, H2799YD, or H3000WC. Also, verify the serial number and system ID listed in the detailed recall information.
Review your system's model number, serial number, and system ID to confirm if it's affected by this recall.
Reach out to GE Healthcare for specific guidance on whether your system is affected and what actions to take.
The recall affects GE Healthcare Infinia II Hawkeye 4 models with specific model numbers: H2799PP, H2799PR, H2799YC, H2799YD, H3000WC, and others as listed in the detailed product information.
The recall is due to the potential for detectors to fall if the system is moved without proper support, which could result in serious injury.
No, but you should check if your system is affected and contact GE Healthcare for specific guidance on next steps.
We’ve drafted a message you can send GE Medical Systems Israel to request your refund or repair — edit it as you like.
Subject: Recall Z-0899-2025 — GE Medical Systems Israel GE Medical Systems Israel Hello, I own a GE Medical Systems Israel that is covered by recall Z-0899-2025 from GE Medical Systems Israel. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.