Intuitive Surgical recalls 8MM Debakey Forceps (model 470036) due to frayed grip cables that may break during use, causing injury. Affected units include specific batch numbers.
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
Intuitive Surgical is recalling 8MM Debakey Forceps with model number 470036 because frayed grip cables could break during use, potentially causing injury. This recall affects specific batch numbers of the product.
Frayed or broken grip cables on these forceps could break during surgery, leading to potential injury from the instrument slipping or causing damage to the surgical site.
Surgical staff and hospitals using the 8MM Debakey Forceps (model 470036) are most at risk. The recall targets specific batch numbers produced between certain dates.
Check for the model number 470036 on the forceps. Affected units have specific batch numbers listed in the recall notice, including K10221128 through K10240801.
Inspect the grip cables for fraying or breaks before use.
The recall is due to increased complaints about frayed or broken grip cables on reusable instruments with jaws.
Affected batch numbers include K10221128, K10221218, K10230219, and other specific numbers listed in the recall notice.
The recall does not require immediate stoppage; however, you should check for frayed cables and avoid use if damaged.
We’ve drafted a message you can send Intuitive Surgical to request your refund or repair — edit it as you like.
Subject: Recall Z-0852-2025 — Intuitive Surgical Intuitive Surgical Hello, I own a Intuitive Surgical that is covered by recall Z-0852-2025 from Intuitive Surgical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.