New Medical Technologies recalls Integra Milt: Cryosolutions Cartridges 4 Pack (Model C-CA-23) due to risk of premature valve emptying during assembly, potentially affecting medical use.
During the assembly of the cartridge to its connecting body part, the valve may open prematurely resulting in partial or complete emptying of the cartridge before its intended use.
New Medical Technologies is recalling Integra Miltex Cryosolutions Cartridges 4 Pack (Model C-CA-23) because the valve may open too early during assembly, causing the cartridge to empty before use. This could lead to improper medical treatment if the cartridge is not fully filled.
The valve in the cartridge can open prematurely during assembly, causing the cartridge to empty partially or completely before it's ready for use. This means the cartridge might not contain enough cryogen (nitrogen) for the intended medical procedure, potentially leading to ineffective treatment.
Medical professionals using these cartridges for cryosurgery are most at risk. Patients receiving treatment with these cartridges may also be affected if the cartridges are not properly assembled.
Check for the model number C-CA-23 and item number 33517 on the cartridge packaging. These cartridges were sold with lot numbers C6741, D0122, D2560, and C6637.
Look for the model number C-CA-23 and item number 33517 on the packaging to confirm if the cartridge is part of the recall.
The valve may open prematurely during assembly, causing the cartridge to empty before intended use, which could lead to ineffective medical treatment.
Check the packaging for the model number C-CA-23 and item number 33517. If it matches, contact New Medical Technologies for further instructions.
The recall is for medical use only, and the risk is low as it affects the cartridge's effectiveness rather than causing injury.
We’ve drafted a message you can send New Medical Technologies to request your refund or repair — edit it as you like.
Subject: Recall Z-1104-2025 — New Medical Technologies New Medical Technologies Hello, I own a New Medical Technologies that is covered by recall Z-1104-2025 from New Medical Technologies. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.