Breathe Technologies recalls Life2000 Ventilators with specific serial numbers due to battery charger alarm malfunction that can disable the ventilator.
There is an issue with the ventilator battery charger that could lead to an audible and visual alarm. If the alarm is engaged, the ventilator becomes inoperable.
Baxter Healthcare Corporation is recalling Life2000 Ventilators with certain serial numbers because the battery charger may trigger an alarm that disables the ventilator. This could be dangerous for patients using the ventilator.
The ventilator battery charger may activate an alarm that disables the ventilator. This could prevent the ventilator from working when needed, potentially harming patients who rely on it.
Patients using the Life2000 Ventilator with the recalled serial numbers are most at risk. Healthcare facilities that own these ventilators should check their inventory.
Check for serial numbers 122010001081 or 00887761978089 (UDI/DI 00887761978089) or 00815410020537. The product was sold with specific model numbers.
If the ventilator alarm is activated, it may become inoperable. Stop using it immediately.
The battery charger may trigger an alarm that disables the ventilator, which could be dangerous for patients.
Check the serial number to see if it's affected. If it is, stop using it and contact Baxter Healthcare Corporation.
Look for serial numbers 122010001081 or 00887761978089 (UDI/DI 00887761978089) or 00815410020537.
We’ve drafted a message you can send Breathe Technologies to request your refund or repair — edit it as you like.
Subject: Recall Z-0887-2025 — Breathe Technologies Breathe Technologies Hello, I own a Breathe Technologies that is covered by recall Z-0887-2025 from Breathe Technologies. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.